Description

The complex, which is based on NIR and Raman spectrometry and may involve additional analytical capabilities, enables real-time collection of data on experiments in the field of wood chemistry, both physico-chemical fractionation and depolymerisation of wood, as well as the modification and polymerisation of wood-based substances and mixtures in water, water-organic, organic and gaseous environments. On the basis of the data collected, process decision-making and control methods that are faster and more economical than the current analytical methods will be developed with the help of reference analyses and chemometrics. The services are offered to research groups engaged in basic research as well as to companies with an applied orientation.

Technical parameters

Parameter Value and unit Additional information
Real-time monitoring of a chemical, biochemical or physical process in the NIR (near-infrared) spectrum region with a transflection probe with a path length of 1 to 10 mm Observable spectrum range:

4000 to 11550 cm-1

Probe path length from 1 to 10 mm

Temperature range up to +130 °C

Working pressure vacuum up to 16 bar

Probe diameter 12 mm

A Bruker MATRIX-F II NIR spectrometer with a FLEX probe is used.

It allows to observe physical and chemical transformations in liquid phase solutions, suspensions, emulsions and to monitor the transportation of substances from one phase to another. With the help of chemometric models, it is possible to create quantitative analysis methods.

Real-time monitoring of a chemical, biochemical or physical process with a 785 nm laser excitation (up to 0.4W) Raman spectroscopy and a 0.2 mm focal length probe Laser excitation wavelength:

785 nm (laser power up to 495 mW)

Observable spectral range:

200 to 3300 cm-1

Probe focal length 0.2 mm

Temperature range -20 to +150 °C

Operating pressure up to 300 bar

Probe diameter 12 mm

PG13.5 reactor interface

 

A Bruker SuperFlux Raman Analyzer with a Tornado TR230 probe is used.

It allows to observe physical and chemical transformations in liquid phase solutions, suspensions, emulsions and to monitor the transportation of substances from one phase to another. With the help of chemometric models, it is possible to create quantitative analysis methods.

Real-time measurement of CO content in chemical and biochemical processes Measurable CO content:

0 to 100 vol% in gas

A Smartgas FlowEVO Plus sensor is used. See valgepealab.com
Collection of reference data necessary for the preparation of real-time process monitoring models using other instrumental analysis methods (IR/Vis/UV spectroscopy, GC, HPLC, MALDI-HRMS, NMR, pH, potentiometry, oven-coulometric KF titration etc.). Qualitative and quantitative analyses starting from simple compounds as water and CO to composition studies of complex mixtures, compilation of reference databases
Modelling, optimisation and validation of the process analysis methodology Possible software packages: Bruker OPUS, S-PACT PEAXACT, etc.

Testing services include:

  • Process monitoring using NIR, Raman, IR, UV/Vis spectrometry
  • Analysis of mixtures using NIR, Raman, IR, UV/Vis spectrometry
  • Real-time analysis of the CO content of the process gas
  • Development, adaptation and validation of analysis methodologies based on NIR and/or Raman spectroscopy

In addition, it is possible to include analyses provided by other laboratories of the Estonian Center of Analytical Chemistry (https://akki.ut.ee/en), the laboratories of the Institute of Chemistry and Physics of the University of Tartu and the Department of Geology.

Example of price offer

  1. Development of the process monitoring NIR/Raman methodology (planning, data collection, data processing, validation, preparation of reports). The cost of a weekly workload is from €5000 + VAT. The necessary chemicals, additional supplies, transportation, and additional analysis services will be added.
  2. NIR/Raman process monitoring consulting (compilation overview of the literature, patents, instrumentation, etc.), data organization, configuration of data collection solution, software configuration, chemometric analysis, preparation of workflow and safety manuals from €120 per hour + VAT

To discuss your potential requirements, contact Ivari Kaljurand at ivari.kaljurand@ut.ee, +372 56626088